Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 31: (1 year) CAL VER /5500 INSTRUMENT: Fluke 33: (1 year) CAL VER /5500 DATE: 08-Nov-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 145 CONFIGURATION: Fluke 5500A STANDARD: Fluke 5500A/COIL ============================================================================= # # Source: # Fluke 31/33 Instruction Manual (PN 896118 Sep. 1991). # Fluke 5500A/COIL Instruction Sheet (PN 600616 Aug. 1995) # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55coil3x.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 ASK- R P F W 1.003 ASK+ K 1.004 HEAD EQUIPMENT SETUP 1.005 HEAD {RMS AMPS TESTS} 1.006 DISP Turn the UUT on. 1.007 PIC 55coil3x 1.008 HEAD RMS AMPS TESTS: {400A Range} 1.009 5500 400.0A 60H SI S T5 1.010 ACC 400.0A 0.50% 0.50U 1.011 MEMI Enter UUT reading in Amps: 1.012 MEME 1.013 MEMC 400 A 8.2U 60H 2.001 5500 200.0A 60H SI S T5 2.002 ACC 200.0A 0.50% 0.50U 2.003 MEMI Enter UUT reading in Amps: 2.004 MEME 2.005 MEMC 400 A 4.2U 60H 3.001 JMPT 14.001 Fluke 31: (1 year) CAL VER /5500 3.002 HEAD {PEAK AMPS TESTS} 3.003 DISP Press the SMOOTH/CREST button twice to select PEAK. 3.004 HEAD PEAK AMPS TESTS: {400A Range} 3.005 5500 565.8Ap 60H SI S T5 3.006 ACC 565.8Ap 0.50% 0.707U 3.007 MEMI Enter UUT reading in Amps peak: 3.008 MEME 3.009 MEMC 400 Ap 17.2U 60H 4.001 5500 282.85Ap 60H SI S T5 4.002 ACC 282.85Ap 0.50% 0.707U 4.003 MEMI Enter UUT reading in Amps peak: 4.004 MEME 4.005 MEMC 400 Ap 8.75U 60H 5.001 5500 141.45Ap 60H SI S T5 5.002 ACC 141.45Ap 0.50% 0.707U 5.003 MEMI Enter UUT reading in Amps peak: 5.004 MEME 5.005 MEMC 400 Ap 4.55U 60H 6.001 5500 113.1Ap 50H SI S T5 6.002 ACC 113.1Ap 0.50% 0.707U 6.003 MEMI Enter UUT reading in Amps peak: 6.004 MEME 6.005 MEMC 400 Ap 3.7U 50H 7.001 HEAD PEAK AMPS TESTS: {40A Range} 7.002 5500 56.6Ap 50H SI S T5 7.003 ACC 56.6Ap 0.50% 0.707U 7.004 MEMI Enter UUT reading in Amps peak: 7.005 MEME 7.006 MEMC 40 Ap 2.0U 50H 8.001 5500 28.285Ap 400H SI S T5 8.002 ACC 28.285Ap 0.52% 0.721U 8.003 MEMI Enter UUT reading in Amps peak: 8.004 MEME 8.005 MEMC 40 Ap 4.55U 400H 9.001 DISP Disconnect the 5500A/COIL from the 5500A AUX Outputs. 9.002 DISP Loop a piece of wire of sufficient gauge from the 9.002 DISP AUX OUTPUT HI to the AUX OUTPUT LO. 9.002 DISP 9.002 DISP Clamp the meter around the loop positioning at the 9.002 DISP alignment marks. 9.003 5500 40 14.145Ap 0.865U 400H SI 2W 10.001 5500 40 11.31Ap 0.64U 60H SI 2W 11.001 5500 40 5.66Ap 0.47U 60H SI 2W 12.001 5500 40 2.825Ap 0.385U 50H SI 2W 13.001 5500 40 1.415Ap 0.345U 50H SI 2W 14.001 HEAD {RMS AMPS TESTS} 14.002 JMPT 14.005 Fluke 31: (1 year) CAL VER /5500 14.003 PIC 55coil3x 14.004 DISP Press the SMOOTH/CREST button to select RMS. 14.005 HEAD RMS AMPS TESTS: {400A Range} 14.006 5500 100.0A 60H SI S T5 14.007 ACC 100.0A 0.50% 0.50U 14.008 MEMI Enter UUT reading in Amps: 14.009 MEME 14.010 MEMC 400 A 2.2U 60H 15.001 5500 80.0A 50H SI S T5 15.002 ACC 80.0A 0.50% 0.50U 15.003 MEMI Enter UUT reading in Amps: 15.004 MEME 15.005 MEMC 400 A 1.8U 50H 16.001 HEAD RMS AMPS TESTS: {40A Range} 16.002 5500 40.0A 50H SI S T5 16.003 ACC 40.0A 0.50% 0.50U 16.004 MEMI Enter UUT reading in Amps: 16.005 MEME 16.006 MEMC 40 A 1.0U 50H 17.001 5500 20.00A 400H SI S T5 17.002 ACC 20.00A 0.51% 0.51U 17.003 MEMI Enter UUT reading in Amps: 17.004 MEME 17.005 MEMC 40 A 1.10U 400H 18.001 DISP Disconnect the 5500A/COIL from the 5500A AUX Outputs. 18.002 DISP Loop a piece of wire of sufficient gauge from the 18.002 DISP 5500A AUX OUTPUT HI to the AUX OUTPUT LO. 18.002 DISP 18.002 DISP Clamp the meter around the loop positioning at the 18.002 DISP alignment marks. 18.003 5500 10.00A 0.70U 400H SI 2W 19.001 5500 8.00A 0.36U 60H SI 2W 20.001 5500 4.00A 0.24U 60H SI 2W 21.001 5500 2.00A 0.24U 50H SI 2W 22.001 5500 1.00A 0.22U 50H SI 2W 23.001 HEAD {FREQUENCY TESTS} 23.002 DISP Press the Hz button. 23.003 HEAD FREQUENCY TESTS: 40A Range 23.004 5500 400.0H 1.1U 10A SI 2W 24.001 5500 60.0H 0.4U 4A SI 2W 25.001 5500 50.0H 0.4U 1A SI 2W 26.001 END